TCID50 Virus Titer Calculator

Calculate virus titer in TCID50/mL using Reed-Muench or Spearman-Karber method.

Usually 1 for 10-fold dilutions
Please check your inputs and try again.

TCID50 Versus PFU

Both measure infectious virus but by different endpoints. TCID50 is a statistical estimate — the dilution infecting half the wells — while PFU counts discrete plaques formed by individual infectious particles.

TCID50PFU
EndpointAny CPE in the wellDiscrete plaque
Requires overlayNoYes
Works for non-lytic virusSometimesOften not
Conversion—PFU ≈ 0.7 × TCID50

The 0.7 conversion comes from Poisson statistics: a well showing CPE may have received one particle or several, and the factor accounts for that. It is an approximation, and virus-specific validation is preferable to assuming it.

Reed–Muench and Spearman–Kärber

Both estimate the 50% endpoint by interpolating between the dilutions bracketing it, since no dilution gives exactly 50% in practice.

MethodApproachNote
Reed–MuenchCumulative totals across dilutionsHistorically dominant
Spearman–KärberWeighted mean of transitionGives confidence intervals

Spearman–Kärber is generally preferred now because it yields a variance estimate, so results can carry error bars. Reed–Muench gives a point estimate only.

Why the Volume Correction Matters

The endpoint dilution gives TCID50 per inoculum volume, not per mL. Dividing by the volume inoculated converts it — with 0.1 mL per well, the titre per mL is ten times the endpoint value.

Omitting this is a common tenfold error. Note also that TCID50 counts only particles capable of causing CPE in that cell line; total particle counts by electron microscopy typically exceed infectious titre by 100 to 1,000-fold, since most virions are defective or fail to establish infection.

Worked Examples

Example 1: SARS-CoV-2: 50% CPE at 10^-5.5 dilution, 0.1mL/well
Titer=10^5.5/0.1
Result: Titer=3.16×10⁶ TCID50/mL
Moderate virus stock
Example 2: Reed-Muench: 80% CPE at 10^-5, 20% at 10^-6
Interp: (80-50)/(80-20)=0.5 steps
Result: 50% dilution at 10^-5.5
Standard interpolation method
Example 3: The volume correction
Endpoint at 10−5.5, 0.1 mL per well
Result: 3.16 × 106 TCID50/mL
The endpoint gives 105.5 per 0.1 mL. Dividing by 0.1 converts to per-mL units.
Example 4: Particle-to-PFU ratio
Electron microscopy count versus infectious titre
Result: 100 to 1,000-fold higher
Most virions fail to establish infection. Infectivity assays measure functional particles, not total virus.

Common Mistakes

⚠️
Forgetting the volume correction

The endpoint gives TCID50 per inoculum volume. With 0.1 mL per well, titre per mL is ten times higher — omitting this is a tenfold error.

⚠️
Treating TCID50 and PFU as interchangeable

PFU is roughly 0.7 times TCID50 by Poisson statistics, but the relationship is virus-specific and should be validated rather than assumed.

⚠️
Assuming infectious titre reflects particle count

Total particles typically exceed infectious units by 100 to 1,000-fold, since many virions are defective.

⚠️
Reading wells too early or too late

CPE develops over days and varies by virus. Scoring before it is complete underestimates titre; overgrowth of controls makes late reading unreliable.

Frequently Asked Questions

TCID50 vs PFU?
TCID50 (tissue culture infectious dose 50%): dilution at which 50% of wells show CPE. PFU (plaque forming unit): visible plaques in agar overlay. Both measure infectious virus. PFU ≈ 0.7×TCID50 (statistically: Poisson). PFU more precise but requires plaque assay.
Reed-Muench vs Spearman-Karber?
Reed-Muench: uses cumulative data, simpler. Spearman-Karber: uses proportions at each dilution, more statistically rigorous. Both give similar results. Karber formula: log TCID50 = last dilution - d(Σp - 0.5) where d=log dilution step, p=fraction positive.
What is the difference between TCID50 and PFU?
TCID50 is a statistical endpoint — the dilution infecting half the wells. PFU counts discrete plaques. PFU is roughly 0.7 times TCID50.
Which endpoint method should I use?
Spearman–Kärber is generally preferred because it provides a variance estimate and therefore confidence intervals. Reed–Muench gives only a point estimate.
Why divide by the inoculum volume?
Because the endpoint dilution gives TCID50 per volume inoculated. Converting to per-mL requires dividing by that volume.
Why is infectious titre so much lower than particle count?
Most virions are defective or fail to establish infection. Particle-to-infectivity ratios of 100 to 1,000 are typical.