Patch Clamp / Ion Channel Calculator
Calculate single-channel conductance, open probability, and current from patch clamp data.
Driving Force, Not Voltage
Single-channel conductance comes from Ohm's law, but the voltage term is the electrochemical driving force — the difference between membrane potential and the reversal potential — not the membrane potential itself.
At the reversal potential the driving force is zero, so no current flows regardless of how many channels are open. This is why Erev identifies which ion a channel carries: it matches the Nernst potential of the permeant species.
| Erev | Ion | Typical channel |
|---|---|---|
| −90 mV | K+ | Inward rectifier, delayed rectifier |
| −65 mV | Cl− | GABAA, glycine receptor |
| 0 mV | Non-selective cation | Nicotinic ACh, NMDA |
| +60 mV | Na+ | Voltage-gated sodium |
A reversal potential near 0 mV is the signature of a non-selective cation channel, since sodium and potassium currents balance there.
Whole-Cell Current
Macroscopic current depends on three separable quantities, and a change in whole-cell current cannot distinguish them. A drug halving the current might halve channel number, halve open probability, or halve single-channel conductance — completely different mechanisms with the same macroscopic result.
Non-stationary noise analysis separates them: plotting current variance against mean current across many trials yields both single-channel conductance and channel number without needing single-channel recording.
| Channel | Conductance |
|---|---|
| BK potassium | 100–250 pS — among the largest |
| Nicotinic ACh receptor | ~30 pS |
| NMDA receptor | ~50 pS |
| AMPA receptor | 5–15 pS |
| CFTR chloride | ~8 pS |
Worked Examples
Common Mistakes
Conductance requires Vm minus Erev. Using Vm alone gives a wrong answer whenever the reversal potential is not zero.
Current is N × Po × i. Halving it could reflect fewer channels, lower open probability or reduced conductance.
Many channels rectify. Inward rectifiers conduct far better in one direction, so a single conductance value does not describe them fully.
It shifts with the concentration gradient. Changing pipette or bath solutions moves the reversal potential predictably, which is how ion selectivity is tested.